- Equipment 6
- gas measurement
- Blood gas
- Polarographic O2 Clarke electrode
- Platinum cathode
- Ag/AgCl anode
- KCl/phosphate electrolyte
- Blood separated by plastic membrane permeable to O2
- 0.7V applied by battery
- CURRENT measured by galvanometer
- can be used for gases AND liquids
- Uses up O2
- may be affected by halothane
- pH electrode
- Hg/HgCl (calomel) reference elecrode
- Ag/AgCl sensing electrode surrounded by KCl buffer encased in pH sensitive glass
- H+ in blood able to pass through H+ sensitive glass bulb, equilibrates with one side in KCl but not the other due to the presence of buffer which creates a potential difference proportional to the H concentration
- No battery, measures VOLTAGE
- Severinghaus CO2 electrode
- modified pH electrode
- KOH and NaHCO3 in solution
- rubber/teflon membrane only allows CO2 to pass into sodium bicarbonate solution, some H+ ions are formed by the Henderson-Hasselbach equation and CO2 can be measured
- ALL
- require regular 2 point calibration
- need to be kept at 37 degrees
- solubility of gases increase w. decreased temp. so hypothermia reduces pO2 and pCO2
- membranes can be damaged
- innaccuracies
- heparin
- acidic, so will cause a spuriously low HCO3- and pH with high K (will actually decrease CO2 due to dilution with the saline)
- delay
- continued metabolism, so low pO2 and pH with high pCO2
- air bubbles
- increase in pO2 and pH, decrease in pCO2
- will cause decrease in pO2 if it is >21kPa
- temperature
- increased solubility of gases with lower temperature so will give falsely high pO2 and pCO2 with low pH if hypothermic patients blood is measured at 37 C
- pH drops with increasing temperature as more H+ ions dissociate
- oxygen
- electrochemical
- polarographic Clarke electrode
- galvanic fuel cell
- gold (Au) cathode
- lead (Pb) anode
- Pb + 2OH- —> PbO + H20 + 2e-
- KOH electrolyte
- No battery, measures VOLTAGE
- 20 second response time
- affected by nitrous oxide (reacts with anode to produce nitrogen)
- physical
- paramagnetic
- water vapour reduces accuracy, theoretical interference from nitric oxide (paramagnetic) and CO2, NO2 and volatiles (diamagnetic)
- due to 2 unpaired electrons in outer shell of oxygen
- types
- new
- test gas and reference gas flow through parallel tubes with a pressure transducer between them, 100Hz pulsing magnetic field applied - the higher the oxygen content the more attracted to the magnetic field and the higher the pressure transduced
- breath by breath, no moving parts, no need for calibration
- old
- 2 nitrogen filled spheres, joined by a bar, suspended by a wire in a magnetic field. Oxygen is attracted to the field and rotates the bar in proportion to its concentration, measured by observing the deflection of a beam of light
- non-deflection
- measures current required to prevent rotation
- may be affected by nitrous oxide and water vapour
- spectroscopy
- oximetry
- CO-oximetry
- measures levels of different types of Hb
- on blood gas machines
- red cells disrupted
- path length fixed
- sapphire glass
- pulse oximetry
- 2 wavelength pulse plethysmograph
- 660 and 940nm (infrared) LEDs
- pulsatile component is approx. 2% of total absorption
- PRO’s
- accurate > 75%
- no issue w. HbF
- 8 CON's
- carboxyHb gives falsely high SpO2 (96%)
- metHb gives falsely low SpO2 (85%)
- with mets you have a LOW chance of living
- also underread with indocyanine green, methyline blue and nail varnish
- dyes and varnish are pretty low end
- interference from other devices
- does not inform about oxygen content or delivery
- inaccurate below 75%
- inaccurate in arrythmias
- Average measurement, so some delay
- can cause burns very rarely!
- laws
- Lambert’s
- the absorption of radiation through a substance increases exponentially as the length of the substance increases
- Beer’s
- the absorption of radiation through a substance increases exponentially as the concentration of the substances increases
- transcutaneous oxygen electrode
- heats skin to 44C
- must be changed every 4 hours
- slow, can cause burns
- 9 oxygen safety features
- Schradar valves
- pin index system
- air
- oxygen
- nitrous
- carbon dioxide
- helium
- heliox
- entonox
- 7
- ‘people in bristol love balloons’ (Severn)
- Cyclopropane
- NIST
- Omeda Link 25 system
- O2 last on the back-bar
- Oxygen knob ridged
- Ritchie whistle
- 7 seconds
- 60 decibels at 1 metre
- sounds when pressure < 200kPa
- galvanic fuel cell with alarm
- should be checked weekly by disconnecting oxygen hose
- paramagnetic analysis with alarm
- pulse oximetry
- vapourisers
- variable bypass
- in circuit
- draw over
- rely on negative pressure
- Goldman
- Epstein and Macintosh of Oxford (EMO)
- heavy, ether only, bellow compensation, water heat sink
- Oxford miniature vaporiser (OMV)
- different volatiles (need to change scale), can be used in series, glycol heat sink, used with tri-service apparatues
- out of circuit
- plenum
- gas pushed through by positive pressure
- Ohmeda TEC (temperature compensated) MK 2, 3, 4, 5
- Aladin Cassette
- maximise surface area for vapourisation with wicks and baffles
- dense metal heat sink
- Boyle’s bottle
- glass: low heat capacity so cooled rapidly and reduced SVP
- improved with water bath
- temperature
- stabilisation
- high heat capacity
- high thermal conductivity
- compensation
- bimetallic strips
- anaeroid bellows
- measured flow
- Gas measurement contd..
- others
- historical
- Engstrom Emma Piezoelectric
- only measure vapours that are soluble in oil (volatiles - cannot distinguish and cannot measure other gases)
- Drager Narcotest
- UV absorption for halothane
- current
- IF absorption spectrophotometry
- IF = 1 - 100 uM
- molecules w. dissimilar atoms absorb IF
- volatiles 3.3uM and 8-9 (better)
- CO2 4.3uM
- N20 4.5uM
- limitations
- collision broadening
- N20 increases CO2, but can be compensated for
- sensitive to water vapour
- sevo, water vapour and nitrous all overlap with CO2
- photoacoustic spectrophotometry
- gas chromatography
- different gases are slowed down different amounts in a carrier gas
- flame ionisation
- thermal conductivity
- beter for inorganic gases
- electron capture
- better for halogenated gases
- ‘(Ciara) O’FITHE’
- very accurate, but expensive, need prior knowledge of likely gases and can’t measure continuously (only for research)
- Raman scattering
- most photons do not change energy state when argon laser is passed through (Raleigh)
- very small number do and specific ones can be detected by a photodetector (Raman)
- very expensive and currently unavailable in UK
- refractrometry
- measures velocity of light in a gas compared to a vacuum
- GOLD STANDARD - Rayleigh (>1m long)
- smaller - Riken
- not used clinically
- can only use for volatiles
- do need calibrating
- influenced by water vapour
- mass spectroscopy
- mass spectrometry
- quadrapole spectrometry
- very accurate
- bulky and expensive
- N20 and CO2 both have MW of 44 so are difficult to measure and one may be over-estimated